SECTION 3—OPERATION
Activating a Mechanical Emergency
Release

Primary Emergency Release

NOTE: For this procedure, refer to FIGURE 3.2.
NOTE: This procedure will bring the boom down if the hand control 
is not functioning properly.
To activate the primary emergency release, insert the tip of a 
pen into the Emergency Down hole () in the control box.

Secondary Emergency Release

NOTE: For this procedure, refer to FIGURE 3.3.
NOTE: All patient lift actuators are equipped with a mechanical 
emergency release. The mechanical release will enable the actuator to 
retract without power.
NOTE: Use the primary emergency release first before using the 
secondary emergency release procedure. This procedure should only 
be used if the primary emergency release procedure is not functioning 
or is unreachable.
NOTE: The lift MUST be under a load for the mechanical release to 
function.
To activate the secondary emergency release, pull up on the 
RED emergency grip and pull down on the boom at the same 
time.

Charging the Battery

Using a Power Cord to Charge the Battery

NOTE: For this procedure, refer to FIGURE 3.2.
NOTE: This procedure only applies to the RPS350‐1FR.
CAUTION
The emergency stop MUST NOT be activated
- otherwise it will be impossible to charge the
battery. While charging takes place, the
patient lift cannot be used. DO NOT attempt
to move the patient lift without unplugging the
power cord from the wall outlet. DO NOT
attempt to use the patient lift if the battery
housing is damaged. Replace a damaged
battery housing before any further use.
1.
Attach the power cord to the control box.
2.
Plug the power cord into a socket outlet.
NOTE: The battery will charge in approximately 4 hours. 
Charging must be done in a room with good air ventilation.
3.
Disconnect the power cord from the socket outlet after the 
battery has been fully charged.
Stand Assist Patient Lift
Control Box
Control Box
Emergency
Up Hole
FIGURE 3.2 Primary Emergency Release
RED Emergency Grip
FIGURE 3.3 Secondary Emergency Release
12
RPS350-1E
Emergency Down
Hole
RPS350-1FR
Emergency Down
Hole
Charger Cable
connects here
Part No. 1078985