Activating a Mechanical Emergency
Release
Primary Emergency Release
NOTE: For this procedure, refer to FIGURE 3.2.
NOTE: This procedure will bring the boom down if the hand control 
is not functioning properly.
To activate the primary emergency release, insert the tip of a 
pen into the Emergency Down hole () in the control box.
Secondary Emergency Release
NOTE: For this procedure, refer to FIGURE 3.3.
NOTE: All patient lift actuators are equipped with a mechanical 
emergency release. The mechanical release will enable the actuator to 
retract without power.
NOTE: Use the primary emergency release first before using the 
secondary emergency release procedure. This procedure should only 
be used if the primary emergency release procedure is not functioning 
or is unreachable.
NOTE: The lift MUST be under a load for the mechanical release to 
function.
To activate the secondary emergency release, pull up on the 
RED emergency grip and pull down on the boom at the same 
time.

Charging the Battery

NOTE: For this procedure, refer to FIGURE 3.4.
NOTE: Invacare recommends the battery be recharged daily to 
prolong battery life. 
NOTE: An audible alarm will sound when battery is low.
1.
Lift UP on the handle on the back of the battery.
2.
Lift the battery up and away from the control box.
CAUTION
Make sure there is an audible click when mount-
ing battery on the battery charger. Otherwise,
injury or damage may occur.
3.
Place the battery on the battery charger. Push the top of the 
battery against the mounting bracket until there is an 
audible click.
NOTE: The charge LED will illuminate. When charged, the LED 
will stop illuminating.
NOTE: It will take approximately four hours to charge a battery that 
requires a full charge.
4.
Lift UP on the handle on the back of the battery.
5.
Lift the battery up and away from the battery charger.
CAUTION
Make sure there is an audible click when mount-
ing battery on the battery charger. Otherwise,
injury or damage may occur.
6.
Place the battery on the control box. Push the top of the 
battery against the mounting bracket until there is an 
audible click.
Part No. 1078984
Control Box
FIGURE 3.2 Primary Emergency Release
RED Emergency Grip
FIGURE 3.3 Primary Emergency Release
Handle
STEPS 1, 2, 4 and 5)
Battery Charger
(STEP 3)
An audible "click"
will be heard when
properly installed
(STEPS 3 and 6)
FIGURE 3.4 Primary Emergency Release
13
SECTION 3—OPERATION
Emergency Down
Hole
Battery
Control Box
(STEP 6)
Stand Up Patient Lift