Activating A Mechanical Emergency Release; Charging The Battery - Invacare JASMINE Owner's Manual

Patient lift
Hide thumbs Also See for JASMINE:
Table of Contents

Advertisement

Available languages

Available languages

Activating a Mechanical Emergency Release

Primary Emergency Release
NOTE: For this procedure, refer to FIGURE 3.2.
NOTE: This procedure will bring the boom down or lift it up if the hand 
control is not functioning properly.
To activate the primary emergency release, insert a pen into the 
hole labeled Emergency Up or Emergency Down on the control 
box.
Secondary Emergency Release
NOTE: For this procedure, refer to FIGURE 3.3.
NOTE: All patient lift actuators are equipped with a mechanical 
emergency release. The mechanical release will enable the actuator to 
retract without power.
NOTE: Use the primary emergency release first before using the 
secondary emergency release procedure. This procedure should only be 
used if the primary emergency release procedure is not functioning or 
is unreachable.
NOTE: The lift MUST be under a load for the mechanical release to 
function.
To activate the secondary emergency release, pull up on the RED 
emergency grip and pull down on the boom at the same time.

Charging the Battery

NOTE: For this procedure, refer to FIGURE 3.4.
NOTE: Invacare recommends the battery be recharged daily to prolong 
battery life. 
NOTE: An audible alarm will sound when battery is low.
1.
Lift up on the handle on the back of the battery.
2.
Lift the battery up and away from the control box.
CAUTION
Make sure there is an audible click when mount-
ing battery on the battery charger to ensure
proper mounting. Otherwise, injury or damage
may occur.
3.
Place the battery on the battery charger. Push the top of the 
battery against the mounting bracket until there is an 
audible click.
NOTE: The charge LED will illuminate. When charged, the LED will 
stop illuminating.
NOTE: It will take approximately four hours to charge a battery that 
requires a full charge.
4.
Lift up on the handle on the back of the battery.
5.
Lift the battery up and away from the battery charger.
CAUTION
Make sure there is an audible click when mount-
ing battery onto the control box to ensure proper
mounting. Otherwise, injury or damage may
occur.
6.
Place the battery on the control box. Push the top of the 
battery against the mounting bracket until there is an 
audible click.
Part No 1150704
Control Box
Emergency Up
Hole
FIGURE 3.2 Primary Emergency Release
FIGURE 3.3 Secondary Emergency Release
Handle
(STEPS 1, 2, 4 and 5)
Battery Charger
(STEP 3)
An audible "click"
will be heard when
properly installed
(STEPS 3 and 6)
FIGURE 3.4 Charging the Battery
13
SECTION 3—OPERATION
Emergency Down
Hole
RED
Emergency Grip
Battery
Control Box
(STEP 6)
Jasmine™ Patient Lift

Advertisement

Table of Contents
loading

Table of Contents