Gauge block
Gauge blocks are available for D30s with ¼" G
or ¼" NPT air connections. To install, ensure
seals are aligned, then use 3 Nm (2.2 lb ft) of
torque when fastening the gauge block to the
positioner using the two screws supplied with
the kit.