Technical specifications
Pressure–flow
curve
60
50
40
30
20
10
0
4
6
Dead space
information
Therapy pressure
Resistance
Environmental
conditions
Operating:
Storage and
Transport:
Gross dimensions
Small:
Mask setting
options
Notes:
•
The mask system does not contain latex, PVC or DEHP materials.
•
The manufacturer reserves the right to change these specifications
without notice.
The mask contains passive venting to protect against
rebreathing. As a result of manufacturing variations, the
vent flow rate may vary.
8
10
12
Mask pressure (cm H
Physical dead space is the empty volume of the mask to
the end of the swivel. Using the small cushion it is 86 mL.
4 to 20 cm H
Drop in pressure measured (nominal)
at 50 L/min: 0.1 cm H
at 100 L/min: 0.7 cm H
Temperature
+41°F to +104°F
(+5°C to +40°C)
-4°F to +140°F
( -20°C to +60°C)
Mask fully assembled - no headgear
5.51" (H) x 3.58" (W) x 3.66" (D)
(140 mm (H) x 91 mm (W) x 93 mm (D))
Select 'STANDARD' as the mask option.
14
16
18
O)
2
O
2
O
2
O
2
Humidity
15% to 95% relative
humidity non-condensing
up to 95% relative humidity
non-condensing
8
Pressure
Flow
(cm H
O)
(L/min)
2
4
19
8
28
12
34
16
40
20
45
20