6.6

Fault finding

Refer to
Table 17
and
18
for fault finding.
Symptom
The Power LED does not illuminate
when the electricity supply is
switched on
The Run LED does not illuminate, or
the pump running status output
signal is open, when the start signal
is set
The pump alarm status output signal
is open and the Alarm LED is off
The pump alarm status output signal
is open and the Alarm LED is
illuminated
A whistling sound is heard as the
pump decelerates
The pump warning status output
signal is open and the warning LED is
illuminated
Other
© Edwards Limited 2007. All rights reserved.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Table 17 - Fault finding
Check
Is the electrical supply correctly
connected?
Is the EPX pump incorrectly connected
to your control equipment?
Is the pump incorrectly connected to
your control equipment?
Has the cooling water supply failed?
Has the run signal been removed and
then re-applied before the pump has
stopped?
Is nitrogen gas flowing?
A419-00-880 Issue F
Action
Ensure that the electrical supply is
correctly connected to the EPX
pump; refer to
Section 3.9
Ensure that the connections
between your control equipment
and the interface connector are
correct: refer to
Section 3.11
Ensure that the connections
between your control equipment
and the interface connector are
correct to
Section
3.11.
Allow the EPX pump to cool down
for at least 20 minutes. Ensure that
the cooling water supply is correctly
connected to the EPX pump, and is
switched on and meets the flow and
pressure requirements of
Section
2.4. Restart the pump.
Removal and re-application of the
run signal may cause a low volume
noise, but it is not harmful to the
pump. It is recommended that the
pump be allowed to come to a halt
before the run signal is re-applied.
Check nitrogen purge gas is flowing
into the gas module. If gas is flowing
and the warning does not clear,
contact your nearest Edwards
service centre.
Contact your nearest Edwards
service centre.
and 3.10.
Page 59