Technical Data; Measurement Characteristics - wtw NiCaVis 705 IQ TS Operating Manual

Optical sensor for nitrat, carbon sum parameters and total suspended solids (tss)
Hide thumbs Also See for NiCaVis 705 IQ TS:
Table of Contents

Advertisement

NiCaVis 705 IQ TS
Measuring principle
Light source
Detector
Wavelengths range
Measuring gap
(optical layer
thickness)
TSS measurement,
5 mm gap width
Nitrate
measurement,
5 mm gap width
Carbon
measurement,
5 mm gap width
ba77189e01
05/2017
8

Technical data

8.1

Measurement characteristics

Spectrophotometric Absorption measurement; integrated microprocessor elec-
tronics, shielded 2-wire connection for power and data transmission.
Lamp type
Photo diodes
200 - 720 nm
NiCaVis 705 IQ TS
Measuring location
Outlet
Measuring location
Outlet
Measuring location
Outlet
*
The UVT-254 value is standardized to 10 mm gap width.
Xenon flashlamp
256
5 mm
Measuring mode
TSS
Measuring mode
NO3
NO3-N
Measuring mode
COD spectr., total
COD spectr., dissolv
TOC spectr., correl.
DOC spectr., correl.
BOD spectr., correl.
SAC 254, total
SAC 254, dissolved
UVT 254, total
*
UVT 254, dissolved
*
Technical data
Measuring range
0.0 ... 900.0 mg/L
Measuring range
0.0 ... 250.0 mg/L
0.00 ... 50.00 mg/L
Measuring range
0.0 ... 800.0 mg/L
0.0 ... 800.0 mg/L
0.0 ... 500.0 mg/L
0.0 ... 500.0 mg/L
0.0 ... 500.0 mg/L
0.0 ... 600.0 1/m
0.0 ... 600.0 1/m
0.0 ... 100.0 %
0.0 ... 100.0 %
Resolution
0.1 mg/L
Resolution
0.1 mg/L
0.01 mg/L
Resolution
0.1 mg/L
0.1 mg/L
0.1 mg/L
0.1 mg/L
0.1 mg/L
0.1 1/m
0.1 1/m
0.1 %
0.1 %
47

Advertisement

Table of Contents
loading

This manual is also suitable for:

Nicavis 705 iq sfNicavis 701 iq niNicavis 705 iq niNitravis 701 iqCarbovis 70 seriesCarbovis 705 iq ... Show all

Table of Contents