Application Characteristics - wtw NiCaVis 705 IQ TS Operating Manual

Optical sensor for nitrat, carbon sum parameters and total suspended solids (tss)
Hide thumbs Also See for NiCaVis 705 IQ TS:
Table of Contents

Advertisement

Technical data
Suitability and areas
of application
Measuring medium
Pressure resistance
Type of protection
Immersion depth
Operating position
48
The quoted measuring ranges are nominal measuring ranges that
are theoretically possible. In practice, real measuring ranges exist
that are given by the limits of photometric determination. The limits
are significantly influenced by the light scattering due to solids and
the absorption of accompanying substances (sample matrix).
Therefore, no fixed measuring range can be determined before-
hand.
8.2

Application characteristics

Municipal wastewater with a low proportion of industrial wastewater
pH value
Flow rate
Temperature
Sensor with connected SACIQ (SW) sensor connection cable:
Max. admissible overpressure
The sensor meets all requirements according to article 3(3) of 97/23/EC ("pres-
sure equipment directive").
Sensor with connected SACIQ (SW) sensor connection cable:
5
IP X8, 1 bar (10
Pa)
min
max
Any
4 - 12
< 3 m/s
0 °C ... + 45 °C
(32 °F ... + 113 °F)
105 Pa (1 bar)
The measuring gap must be
filled with test sample.
10 m depth
NiCaVis 705 IQ TS
ba77189e01
05/2017

Advertisement

Table of Contents
loading

This manual is also suitable for:

Nicavis 705 iq sfNicavis 701 iq niNicavis 705 iq niNitravis 701 iqCarbovis 70 seriesCarbovis 705 iq ... Show all

Table of Contents