General Steps To Be Taken - wtw NiCaVis 705 IQ TS Operating Manual

Optical sensor for nitrat, carbon sum parameters and total suspended solids (tss)
Hide thumbs Also See for NiCaVis 705 IQ TS:
Table of Contents

Advertisement

NiCaVis 705 IQ TS
Example 1
Example 2
ba77189e01
05/2017

5.2.2 General steps to be taken

Depending on the application site and the level of contamination of the sensor
as well as the coming job, the cleaning procedure includes the following parts:
 Every cleaning procedure starts with a basic cleaning. It removes tough
grime such as incrustation of fouling matter, algae and biological deposits. If
the sensor surface might be contaminated with pathogenic germs it should
also be disinfected.
 Cleaning the measuring gap: Removal of lime deposits.
 Cleaning the measuring gap: Removal of grease deposits.
A heavily contaminated sensor, for which a
should be carried out, must run through all cleaning steps properly. During the
cleaning process always keep all surfaces moist to prevent already dissolved
contamination and detergents from drying ("working wet-on-wet"). Finally, thor-
oughly rinse off all remaining cleaning agents with ultrapure water.
If the sensor from example 1 should be cleaned on site during a routine visual
inspection and should then continue measuring, a basic cleaning might possibly
be sufficient.
The quoted examples and the following descriptions are intended to
give a general guideline. Due to the sensor's wide range of applica-
tion, the type and level of contamination can considerably vary.
Therefore, the cleaning procedure has to be individually composed.
If necessary, individual cleaning steps have to be changed or
repeated several times until the result meets the requirements.
Maintenance and cleaning
Sensor check
Zero adjustment
or