Maintenance and cleaning
Cleaning agents
Accessories
38
5
Maintenance and cleaning
5.1
Maintenance
The UV-VIS- NiCaVis 705 IQ TS sensor operates maintenance-free.
5.2
Sensor cleaning
5.2.1 Cleaning agents and accessories
To clean the sensor, use the following cleaning agents only:
Contamination
Water-soluble substances
Lime deposits
Grease deposits
* For the final cleaning prior to a
cleaning agents with purity grade "for analysis". Finally, thoroughly rinse the
sensor with ultrapure water.
Sponge
Plastic-bristled brushes of different sizes.
Interdental brushes are suitable for the cleaning of the compressed air chan-
nels.
Flocked cleaning cards (see accessories)
Lens tissues
Wash bottle
NOTE
Never use any pressure washer, steam blaster, abrasives or metal brushes to
clean the sensor. Risk of damaging the sensor.
Cleaning agents
– Primary cleaning: Tapwater, warm if
possible
(30 - 50 °C)
– Precision cleaning: Household wash-
ing-up liquid + ultrapure water.
5 % hydrochloric acid *
Mixture of isopropanol and water (approx.
70 % isopropanol) *
Sensor check
Zero adjustment
or
NiCaVis 705 IQ TS
, only use
ba77189e01
05/2017