11. Maintenance Tips

To secure the accuracy of your measuring instrument we recom-
mend the annual check by an authorized ecom partner. In the case
of strong demand (e.g. permanent several hours of measurement
per day, rough conditions etc.) shorter intervals between checks
should be selected - please contact your ecom partner. All ecom
partners are listed under www.ecom.de. Do not use other sensor
than supposed by our factory and be aware that service done by
service stations not authorised by ecom will lead to expiration of any
warranty.
The following tips for the daily care of accessible parts and compo-
nents should be helpful:
Condensate pad
Condensate pad
Regularly check the condition of the condensate pad. It should be
changed:
- If it is soiled darker than no. 3 on the soot scale
- If its extension is on mark "MAX"
Replace the condensate pad as follows:
1. Remove hose from the cap and unscrew the cap
2. Unscrew the retaining bolt with spring
3. Replace old condensate pad against new
4. Scew on retain bolt with spring again
5. Scew on cap and replace the hose to the cap
ecom-B
Page 39