Call up the menu „Pressure Tests" with the pre-programmed
measurement routines „Pressure Test", „Loading Test", „Tight-
ness Test" and „Usage property". Parameters to each measure-
ment routine can be adjusted in the menu „Setup".
Setup
Scroll with cursor keys <Up/Down> until the menu "Setup". Press
<Enter> to activate the menu. The following parameters can be ad-
justed for the corresponding measurement routine:
Pressure Test
- Stabilization time (1 - 10 min, default: 1 min)
- Measurement time (10 - 120 min, default: 5 min)
Loading Test
- Stabilization time (1 - 10 min, default: 1 min)
- Measurement time (10 - 120 min, default: 10 min)
- Test pressure (900 - 1200 hPa, default: 1000 hPa)
Tightness Test
- Stabilization time (1 - 10 min, default: 1 min)
- Measurement time (10 - 120 min, default: 10 min)
- Test pressure (90 - 160 hPa, default: 150 hPa)
Usage property
- Stabilization time (1 - 270 min, default: 1 min)
- Measurement time (10 - 240 min, default: 10 min)
- Test pressure (20 - 160 hPa, default: 50 hPa)
- Max. operation pressure (10 - 100 hPa, default: 23 hPa)
Each parameter can be adjusted the same way:
1. Select the parameter with cursor keys <Up/Down>.
2. Press <Enter> to activate.
3. select the digit with the keys <Store> (left), <Print> (right) and
adjust with the cursor keys <up/down>.
4. Confirm with <Enter>.