Pressure Test
The "Pressure Test" up to 100 hPa is deposited as a measurement
routine in the ecom-B. Proceed as follows:
1. Close the conduit with a suitable adapter (test stopple, high-
pressure stopple or one-pipe counter cap).
2. Connect the components as described before.
3. Scroll with cursor keys <Up/Down> to the menu "Pressure Test".
Activate with <Enter>.
4. Create the pre-adjusted test pressure (max. 100 hPa) with the air
pump.
5. Interrupt the connection to the air pump (switch-off the ball valve)
and start "Pressure Test" with <Enter>.
6. Wait for stabilization time (the measurement will start automatical-
ly).
7. Once the measurement time is over, the result is displayed and
can be printed (IR printer required) by pressing <Print>.
8. If the menu "Pressure Test" is selected again, so the result can
be called up with "No" / <Enter> (as long as the ecom-B is on) or a
new measurement can be started with "Yes" / <Enter> (switching
from "No" to "Yes" with cursor keys <Up/Down>).