Bath Pre-Set; Specification - ZOLL thermogard xp Operation Manual

Hide thumbs Also See for thermogard xp:
Table of Contents

Advertisement

WARNING! Change to Initial Specification of System.
In the initial release of the System there were no settable patient temperature
alarms. There were fixed high and low patient temperature alarms. These
fixed alarms have been made programmable.
The patient temperature alarms MAY NOT be deactived. They may be muted
for a 2 minute period.
NOTE: Patient Temperature Alarms in STANDBY
If the System is in STANDBY and the T1 temperature probe is connected to
the System but NOT inserted into the patient you will experience "Lo" patient
temperature alarms if the ambient temperature of the exposed probe is below
the alarm limit. To avoid the annoyance of the alarm under these conditions,
simply unplug the T1 temperature probe from the System until you are ready to
connect to the patient. You may not enter RUN without the probe connected
to the System.

Bath Pre-set

The menu allows the user to select from either: Pre-Cool, Pre-Warm or None.
This selection operates only in STANDBY and is cancelled when the system
enters RUN.
Figure 4-13. Bath Pre-Set Menu
17
Operation
600479-001 Rev 2
48/147