ZOLL THERMOGARD HQ Instructions For Use
ZOLL THERMOGARD HQ Instructions For Use

ZOLL THERMOGARD HQ Instructions For Use

Surface start-up kit

Advertisement

Surface Start-Up Kit
Instructions for Use
106875-002 Rev. 2 01/2024

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the THERMOGARD HQ and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for ZOLL THERMOGARD HQ

  • Page 1 Surface Start-Up Kit Instructions for Use 106875-002 Rev. 2 01/2024...
  • Page 2: Operating Life

    Focus attention on all bony prominences. • When a Thermogard HQ console is used with a Thermogard HQ Surface Start-Up Kit and STx Vest, the Treatment Mode is set to Fever Control. The console does not warm the patient when the Fever Control mode is selected. The Lo patient temperature alarm limit ensures that an alarm occurs should the patient stop regulating his/her own body temperature.
  • Page 3 ZOLL Thermogard HQ™ Surface Start-Up Kit 2. Set Up Tubing 2.0 FILL RESERVOIR BAG A. Fill the reservoir bag with water and fasten cap. B. Hang the reservoir bag on console hook. 2.1 INSERT COIL A. Open coolant well cap.
  • Page 4: Troubleshooting

    8. Pull the air trap up and out of the chamber. 9. Remove the water reservoir from the hook. 10. Bundle up the STx Vest and the collection of Thermogard HQ Surface Start Up Kit components, still connected together, and gently deposit them in the medical waste container.
  • Page 5: Catalog Number

    Copyright ©2023 ZOLL Medical Corporation. All rights reserved. Thermogard, Thermogard HQ, Temperature Management System, STx Vest, and IVTM are trademarks or registered trademarks of ZOLL Medical Corporation and/or its subsidiaries in the United States and/or other countries. All trademarks are the property of their respective owners.