ZOLL thermogard xp Operation Manual page 123

Hide thumbs Also See for thermogard xp:
Table of Contents

Advertisement

Alarm Message
DOWNLOAD ERROR.
PLEASE CHECK
EXTERNAL
COMPUTER
PATIENT DATA LOG
FULL. PLEASE
DOWNLOAD NOW
Hi PATIENT
TEMPERATURE
ALARM
Lo PATIENT
TEMPERATURE
ALARM
PRIMARY
TEMPERATURE
PROBE
DISCONNECTED*
PRIMARY
TEMPERATURE
PROBE DISLODGED*
PUMP DOOR OPEN*
PUMP FAILURE*
Cause
Patient trend data download
has failed or there is no
computer connected to the
System.
Patient trend data log is full.
Only two hours of space
remain.
The patient's temperature is
above the programmed Hi
patient temperature alarm
value.
The patient's temperature is
below the programmed Lo
patient temperature alarm
value.
The primary patient
temperature probe is
disconnected.
Temperature output from the
primary patient temperature
probe changed by –0.2° C or
more within 10 seconds.
The pump cover is open.
The pump has failed.
4
Alarms & Corrective Actions
Corrective Action
Check the connection
between the computer and the
System and select "Try again."
If problem persists, select
"Cancel" and contact ZOLL
technical support.
Download patient trend data.
If data is not downloaded, it
will be overwritten.
The System alarm will reset
when the patient's
temperature falls below the
programmed Hi patient
temperature alarm value.
The System alarm will reset
when the patient's
temperature rises above the
programmed Lo patient
temperature alarm value.
Verify position of primary
temperature probe. Assure
that primary probe is plugged
into socket T1, or if an
interconnect cable is used, the
probe is plugged into the cable
and the cable is plugged into
T1.
Replace probe in patient.
Close the cover.
Inspect the pump. Clear any
obvious faults and restart. If
the problem persists,
discontinue use and contact
your ZOLL service
representative (see Chapter
9).
600479-001 Rev 2
123/147