the flow indicator does not rotate freely during patient treatment, inspect the
entire tubing circuit for kinks or other restrictions to flow. Tapping on the flow
indicator will liberate any trapped air bubbles and return them to the saline bag.
Figure 4-53. Inline Flow Indicator.
WARNING! INVESTIGATE AIR TRAP ALARMS.
If an "Air Trap Fault" alarm occurs, it is likely that there is a leak in the tubing
circuit or the catheter has failed. Refer to Chapter 6 – Alarms and Corrective
Actions for assistance. Do not keep replenishing saline that is being rapidly
depleted–a problem exists that must be immediately remedied.
WARNING! START-UP KIT LIFETIME IS SEVEN (7) DAYS.
The designed operating lifetime for Start-up Kit components is seven (7) days
of continuous operation. If a patient must be treated for a longer period, a new
Start-up Kit must be installed in the System. Failure to adhere to this time limit
may cause injury to the patient.
44
Operation
600479-001 Rev 2
75/147