The following electrical and operating characteristics are not
specifications, but are typical operating characteristics for the Keysight
16860 series logic analyzers.
This product is designed for use in Installation Category II and Pollution Degree 2
environment.
This instrument has auto-ranging line voltage input. Ensure that the supply voltage
is within the specified range and voltage fluctuations do not to exceed 10 percent of
the nominal supply voltage.
Do not apply more than 5.5 volts to Trigger In BNC or Trigger Out BNC.
Electrical Characteristics
Power Requirements
16861A, 16862A, 16863A, and 16864A: 00-120VAC 50/60/400Hz &100-240VAC
50/60Hz mains capable of supplying 325 watts.
CAT II (Line voltage in appliance and to wall outlet).
Installation category II, Pollution degree 2.
Environmental Characteristics (Operating)
Temperature
Instrument: 5° C to 40° C
Altitude
4000 m
Humidity
Maximum 90% Relative humidity, non-condensing
Intended for use in an indoor lab environment