The different menu points in the OEM-Menu are listed in the
Table 3 below.
Table 3. OEM Menu Options
Menu
Description
OEM MENU
OEM Menu Prompt
Alarm set point for maximum
Max Temp
allowable temperature.
Time setting for how long the
Recirculation
Smart2Pure Pro will
Interval
recirculate water in the
Dispense Mode
When the unit switches from
standby mode into dispense
Rinse Interval
mode (or tank filling) a short
rinse of the RO cartridge
occurs
The date and time can be set
Date & Time
in this screen.
If printer option is used, this
menu is where to set the
Print Interval
frequency of the data
printout
Setting for different
Language
languages
Allows for water purity to be
μS/cm -
read in conductivity (μS/cm)
MΩ*cm
or resistivity (MΩ*cm)
Allow purity on the display to
Temp. comp.
be temperature
compensated to 25°C
Tank level settings (OFF/ON)
to adjust when filling of the
tank stops (OFF) and when it
OFF/ON/
starts again (ON). Also, sets
Empty
the level at which the
recirculation pump can turn
on (Empty)
Turns on leak detection and
threshold. If ON, if leak is
Leak
detected all pumps and feed-
Detection
water solenoid valves are
de-energized.
Smart2Pure Pro Ultrapure water system
Settable
Factory
Range
Setting
N/A
N/A
1 - 50°C
40°C
10, 20, 30
10 minutes
minutes
0.1 - 10
2 seconds
seconds
1-12 month,
1-31 days,
N/A
0-24 hours,
0-60minutes
5 minutes - 12
5 minutes
hours
English,
French,
English
German
μS/cm -
MΩ*cm
MΩ*cm
ON - OFF
off
Off: 40 - 100%
Off: 100%
On: 10-70%
On: 70%
Empty: 10-
Empty: 25%
25%
ON - OFF
On
Actions
Press ENTER key to enter into the
OEM Menu
Use Up/Down/Right Arrow keys to
adjust setting and press ENTER to
confirm Change
Use Up/Down/Right Arrow keys to
adjust setting and press ENTER to
confirm Change
Use Up/Down/Right Arrow keys to
adjust setting and press ENTER to
confirm Change
Use Up/Down/Right Arrow keys to
adjust setting and press ENTER to
confirm Change
Use Up/Down/Right Arrow keys to
adjust setting and press ENTER to
confirm Change
Use Up/Down/Right Arrow keys to
adjust setting and press ENTER to
confirm Change
Use Up/Down Arrow keys to adjust
setting and press ENTER to confirm
Change
Use Up/Down Arrow keys to adjust
setting and press ENTER to confirm
Change
Use Up/Down/Right Arrow keys to
adjust setting and press ENTER to
confirm Change
Use Up/Down Arrow keys to adjust
setting and press ENTER to confirm
Change