Screws
Figure 8. Upper mount bracket
5. Screw the lid on top of the tank.
6. Screw in the two screws with a Phillips screw driver into
the wall where you have put in the plugs before (refer to
Figure 8).
7. After mounting the complete wall mounting bracket,
connect the tubing and level control cable as described
in the section Connections of the Smart2Pure Pro
Pure Water Storage Tank.
Figure 9. Wall mounted 30 L tank
Smart2Pure Pro Ultrapure water system
Attaching Wall Mounting
Bracket for 60 L Pure Water
Clearance Holes
Tank
1. Locate the two metal brackets and the baseplate. Screw
in the six screws and hexagon nuts supplied with the
brackets as shown in the diagram.
Figure 10. Assembling 60 L mounting bracket
2. Hold the wall mounting bracket at the desired position on
the wall and mark the six boreholes for fixing the wall
mounting bracket. Use an 8 mm or 0.31 inch twist drill to
make the holes and put in the six S8 plugs which are
supplied with the wall mounting bracket (refer to
Figure 11).
3. Attach the wall mounting bracket to the wall by screwing
in the six supplied screws with a Phillips screw driver.
Figure 11. Mounting bracket to wall
4. Place the 60 L pure water tank with two people on the
wall mounting bracket.
Screw
Hexagon
Nuts
Metal Bracket
The Installation Area |
Baseplate
Hole Marking