Thermo Scientific Smart2Pure Pro Installation And Operation Manual page 18

Barnstead water system
Table of Contents

Advertisement

Water connections
Feed-water tubing
Concentrate tubing
Tank Flow tubing (Smart2Pure Pro)
Tank back flow tubing (Smart2Pure Pro)
Remote dispenser tube jumper
External dispensing valve
Electrical connections
Input voltage
Output voltage
System connection
Serial Interface
Potential free contact
Protection class
Ambient conditions
Operation area
Maximum altitude above sea level
Temperature range
Line-voltage variation
Transient over voltages
Ventilation requirements
Degree of pollution
Airborne sound emission
Sound pressure level
Smart2Pure Pro Ultrapure water system
AC 100 – 240 VAC, 50 – 60 Hz, 2.0 A Max
DC 24 V, 5.0 A Max
DC 24 V, 120 W
RS232
Maximum 30 V, 2 A
Class II (external SMPS certified as Class I)
During Operation
Indoor rooms
up to 2000 m
Min. +2°C, Max +40°C, 80% rel. rH,
non-condensing
Not more than +/- 10% of the line
voltage
As usually occur in the supply network
(overvoltage category II acc. to IEC
60364-4-443)
Note: The rated level of transient
overvoltage is the withstand impulse
voltage acc. to overvoltage category II
of IEC 60364-4-443.
no special requirements
2
Tube ¼" outer diameter
Tube ¼" outer diameter
Tube 6 mm outer diameter
Tube 6 mm outer diameter
Tube ¼" outer diameter
Tube ¼" outer diameter
Storage
Indoor rooms
up to 2000 m
Min. +2°C, Max. +60°C, 90% rel. rH,
non condensing
-- (not applicable)
-- (not applicable)
no special requirements
2
49 dB (A)