Thermo Scientific Smart2Pure Pro Installation And Operation Manual page 28

Barnstead water system
Table of Contents

Advertisement

Figure 25. Sterile overflow tube
Install CO
Vent Filter
2
Remove plug at top of tank and thread CO
place.
CO
Vent Filter
2
Figure 26. CO
Vent Filter installation
2
Mounting the External
Dispensing Valve to the
Smart2Pure Pro System
(Optional)
Note: An external dispensing valve with a t-connector is
provided with the assembly kit to dispensed pure water ASTM
Type II quality from the internal or external tank from the
Smart2Pure system.
Smart2Pure Pro Ultrapure water system
Follow the steps below to mount the external dispensing
valve:
1. Switch the system OFF. (unplug the main power supply)
2. a
Tank
Tank Overflow
Fitting
AD 8 mm / 0.31"
Tube
Figure 27. Tube jumper
vent filter in it's
2
Installed
CO
Vent Filter
2
Figure 28. T-Connection and external dispensing
Locate the external dispensing valve, the t-adapter
and the outer diameter ¼" tube.
b. Remove the tube jumper from the connection panel
of the system.
Tuber Jumper
Dispenser
Back Flow
c. Cut from the outer diameter ¼" tube two tubes with a
length of 80 mm/ 3.2 inch and attach the tubes into
the t-adapter and the connection panel. Attach the
external dispensing valve with a third outer diameter
¼" tube to the t-adapter.
Connection Panel
External
Dispensing Valve
Outer Diameter
¼" tube
valve
Dispenser
Flow
T-Adapter
Outer Diameter
¼" tube
Installation |