5.11 Memory management

5.11.1 Storing a reading

There are 100 files (F00-F99) which can be used to store the measurement values
inside the instrument. At most 100 records can be stored in each file.
The following steps outline how to do this:
1) The
key has to be pressed to activate the
【File Name】item on the main measurement screen.
2) The
and
key are to be used to select the desired file to save the data.
3) After a new measurement reading appears, the
the measurement value to the current file. If the 【Auto Save 】function is activated,
the measurement value will be automatically saved to the current file after a new meas-
urement operation.
5.11.2 Viewing a stored reading
key has to be pressed multiple times until the 【file name 】area on the meas-
The
urement screen is highlighted. The
number.
delete the marked file
delet all files
oder
select the file to store
escape the menu
Press the
key until {Record Count} is shown at the display. The
to be used to change the record count.
delete the highlited measurement value
delete all measurement values
oder
escape the manu
TU_US-BA-e-1714
key has to be pressed to save
and
key are to be used to change the record
and
key are
17