Calibrate objective
magnification
Unassign objectives
52
1.
Go to the Settings  Objectives, then click the Calibrate icon next to the
newly installed objective to launch the Objective Calibration tool.
2.
Mount the EVOS
Calibration Slide in the vessel holder, select the objective
you want to calibrate, then click Calibrate Objective.
3.
Follow the onscreen instructions to calibrate the objective magnification for the
selected objective. For more information, see "Calibrate the objectives", page 62.
After removing an objective from the objective turret, unassign the objective from
the corrsponding turret position.
1.
Go to the Settings  Objectives, and click the Unassign objective button
that corresponds to the turret position from which you have removed
the objective.
EVOS
M5000 User Guide