Technical specifications

Operation temperature/
humidity:
Transportation/ Storage
temperature/humidity:
Weight:
Dimensions:
Measuring method:
Measurement range:
Cuff pressure display:
Voltage source:
40
10 to 40 °C (50 to 104 °F), 15 - 90 %
relative maximum humidity at an
atmospheric pressure range of 860hPa
to 1060 hPa.
-20 to 55 °C (-4 to 131 °F)
15 - 90 % relative maximum humidity
1100 g (including rechargeable battery
pack)
200 x 125 x 90 mm
Oscillometric, corresponding to
Korotkoff
30 - 280 mmHg – blood pressure
40 - 200 beats per minute – pulse
Range: 0 - 299 mmHg
Resolution: 1 mmHg
Static accuracy: pressure within ± 3
mmHg
Pulse accuracy: ±5 % of the readout
value
Rechargeable battery pack;
4.8V C3500 mAh
Mains adapter DC 7.5V, 2 A
Reference to Standards: • Device corresponds to the requirements
Electromagnetic
compatibility:
0044
Type BF applied part
Manufacturer
Class II equipment
For indoor use
Microlife reserves the right to alter technical specifications without prior written notice.
of the standard for non-invasive blood
pressure monitor.
EN 1060-1
EN 1060-3
EN 1060-4
IEC 60601-1
IEC 60601-1-2
• Device fulfills the stipulations of the
standard IEC 60601-1-2.
The stipulations of the EU Directive 93/42/
EEC for Medical Devices Class IIa have been
fulfilled.