5 Care and cleaning instructions for the practice team
5.3 Care, cleaning agents, and disinfectants
93°C
200°F
42
5.3

Care, cleaning agents, and disinfectants

NOTICE
Authorized care, cleaning agents, and disinfectants
Use only care, cleaning and disinfecting agents approved by the
manufacturer
A continuously updated list of approved agents can be downloaded from
the Internet at http://www.sirona.com/manuals. Click on the
documents"
menu item and then open the
disinfection agents"
If you do not have access to the Internet, please contact your dental depot
to order the list (REF 59 70 905).
5.4

General processing instructions

The general processing instructions apply to the operating light and
ceiling model unless there are any other additional product-specific
processing instructions in these operating instructions. The
manufacturer's instructions related to disinfectants must be observed
(temperature, concentration, exposure times, etc.).
Manual cleaning
The equipment can be cleaned manually using a cloth or soft brush.
Unless specified otherwise, use drinking water that is warm to the touch
to clean surface contamination.
Mechanical cleaning
Thermal disinfection at up to 93°C in accordance with ISO 15883-1 is
possible with labeled components. Use an alkaline cleaning agent for
this. A laboratory dishwasher can be used as cleaning equipment (e.g.
Miele G7781). The setting for the cleaning equipment as well as the type
and quantity of the cleaning agent can be found in the operating
instructions for the relevant device.
Manual disinfection
The operating light and ceiling model can be wiped with surface
disinfectants. Use a soft colorless cloth and approved disinfectant for
disinfection. Unless otherwise stated in the operating instructions, other
disinfection procedures should not be applied (spray disinfection,
immersion bath, etc.). When using disinfectants please follow the
information in their usage instructions (temperature, concentration,
exposure times, etc.).
Mechanical disinfection
Thermal disinfection at up to 93°C in accordance with ISO 15883-1 is
possible with labeled components. Use an alkaline cleaning agent for
this. A laboratory dishwasher can be used as disinfection equipment (e.g.
Miele G7781). The setting for the disinfection equipment as well as the
type and quantity of the disinfectant can be found in the operating
instructions for the relevant device.
Operating Instructions LEDview Plus
"Care, cleaning and
document.
Dentsply Sirona
"General
65 67 510 D3642
D3642.201.01.03.02 01.2018