Handling Ph And Ion-Selective Electrode Sensors (Ises); Using The Rdo Sensor And Rdo Fast Cap - In-situ Aqua TROLL 500 Operator's Manual

Multiparameter sonde
Hide thumbs Also See for Aqua TROLL 500:
Table of Contents

Advertisement

Handling pH and Ion-Selective Electrode Sensors (ISEs)

!
Salt may accumulate around the reference junctions of the ammonium, chloride, nitrate and pH
sensors. Rinse with deionized water to remove any buildup.
!
Before using the pH or ISE sensors for the first time, replace the reference filling solution. Condition
the sensors by soaking in calibration standard for 4-24 hours prior to deployment.
!
The wiper can severely reduce the life of the RDO Fast Cap. Wear will vary by application. Verify sensor
performance prior to use and replace the Fast Cap if damaged.
RDO Fast Cap
1-970-498-1500

Using the RDO Sensor and RDO Fast Cap

Potential salt buildup. Rinse with deionized
water if necessary.
1
Unscrew reference junction.
2
Replace reference filling solution.
Inspect foil prior to each use. If
damaged, replace cap.
Wiper
17
www.in-situ.com

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents