Sample Result (Turbidity/Suspended Solids Sensors); Dissolved Oxygen Sensor - 100% Open Air Calibration - In-situ Partech 228531 Operator's Manual

750w2 monitor c/w carrying bag and usb cable
Table of Contents

Advertisement

Sample Result (Turbidity/Suspended Solids Sensors)

Use Sample Result to adjust the value stored earlier with the Take Sample function.
You must know the value of the calibration solution based on a lab analysis.
Press Menu on the
keypad to display the
main menu.
The sample result
screen displays sensor
info. Press OK.

Dissolved Oxygen Sensor - 100% Open Air Calibration

Calibrate the RDO sensor using the saturation method described below. See the
manual for the Partech 750w2 Monitor to get more information about calibration.
Press Menu on the
keypad to display the
main menu.
The monitor displays
sensor info. Press OK.
Select Measurement
Config with the arrow
keys. Press OK.
The screen shows the
date and time of the
last sample. Press OK.
Select Measurement
Config and press the
OK key.
Calibrate the sensor
in open air free from
drafts. Press OK.
Choose Sample
Result and press OK.
Use the arrow keys to
adjust the value. Press
OK.
Choose DO Saturation
and press OK.
Wait for a saturation
percentage of roughly
90%. Then press OK.
7
Use the arrow keys
to select Suspended
Solids. Press OK.
The screen displays
a success message.
Press OK.
Highlight Calibrate
and press OK.