In-situ Aqua TROLL 500 Instruction Sheet

In-situ Aqua TROLL 500 Instruction Sheet

Bga-pc sensor
Hide thumbs Also See for Aqua TROLL 500:

Advertisement

Install sensor.
1
Rinse sensor with clean
water before use.
Install sensor. Do not
twist.
Calibrate and deploy.
2
!
For detailed calibration instructions, see the instruction manual or quick start guide for your In-
Situ instrument.
Connect to the
instrument with VuSitu
or Win-Situ software.
Page 1

Aqua TROLL 500/600 BGA-PC Sensor Overview

The In-Situ blue-green algae/phycocyanin sensor measures BGA-PC levels in
natural water, surface water, groundwater, produced water and aquaculture
applications.

Getting Started

Remove restrictor from
the instrument.
Place restrictor in
calibration mode.
Select Calibrations from
the menu.
970-498-1500
Instruction Sheet
Remove sensor port plug
if installed. Do not twist.
Choose the BGA-PC
option and follow the
instructions.
www.in-situ.com
Lubricate o-ring at
bottom of sensor.
Flip the restrictor into
deployment mode after
calibration
2019-01-9 | 0038922

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Aqua TROLL 500 and is the answer not in the manual?

Questions and answers

Summary of Contents for In-situ Aqua TROLL 500

  • Page 1: Getting Started

    Instruction Sheet Aqua TROLL 500/600 BGA-PC Sensor Overview The In-Situ blue-green algae/phycocyanin sensor measures BGA-PC levels in natural water, surface water, groundwater, produced water and aquaculture applications. Getting Started Install sensor. Rinse sensor with clean Remove restrictor from Remove sensor port plug Lubricate o-ring at water before use.
  • Page 2 Expected RFU Value at 25° C Concentration 25° C 0 µg/L (deionized none water) 100 µg/L 1.0 mL 80 µg/L (ppb) 200 µg/L 2.0 mL 159 µg/L (ppb) 400 µg/L 4.0 mL 309 µg/L (ppb) Page 2 970-498-1500 www.in-situ.com 2019-01-9 | 0038922...

This manual is also suitable for:

Aqua troll 600

Table of Contents