You must have a Temperature or Conductivity/Temperature sensor installed to use the
pH/ORP sensor.
Replace the reference filling solution.
1
Before calibration and deployment, replace the filling solution according to the instructions
below. Repeat filling procedure between deployments or when measurements begin to drift.
Remove sensor from
sonde and unscrew
reference junction.
Squeeze a steady stream
of solution into the
reservoir while slowly
pulling out the tube.
Page 1
Aqua TROLL 500/600/700/800 pH/ORP Sensor
The In-Situ pH sensor measures pH levels in natural water, surface water,
groundwater, produced water and aquaculture applications.
Getting Started (3 steps)
Don't pour solution
down the drain. Pour it
onto a paper towel and
discard.
Overfill slightly. Reinstall
reference junction cap
and tighten until it
touches sensor body.
970-498-1500
pH/ORP Sensor
Lightly shake the bottle
of reference filling
solution to mix.
Turn the cap 90° more
(one quarter of a turn) to
secure.
www.in-situ.com
Insert the fill tube into
the bottom of reservoir.
2024-02-26 | 0063472

Advertisement

loading
Need help?

Need help?

Do you have a question about the Aqua TROLL 500 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for In-situ Aqua TROLL 500

  • Page 1 Sensor Aqua TROLL 500/600/700/800 pH/ORP Sensor The In-Situ pH sensor measures pH levels in natural water, surface water, groundwater, produced water and aquaculture applications. You must have a Temperature or Conductivity/Temperature sensor installed to use the pH/ORP sensor. Getting Started (3 steps) Replace the reference filling solution.
  • Page 2 Connect to the deployment mode after instrument with VuSitu. the menu. option and follow the calibration instructions. For detailed calibration instructions, see the instruction manual or quick start guide for your In-Situ instrument. Page 2 970-498-1500 www.in-situ.com 2024-02-26 | 0063472...
  • Page 3 Storage Solution or pH 4 storage cap. calibration standard. Salt crystals may form on the sensor during storage. These are normal and will not interfere with sensor performance. Page 3 970-498-1500 www.in-situ.com 2024-02-26 | 0063472...

This manual is also suitable for:

Aqua troll 600Aqua troll 700Aqua troll 800