Replace the reference filling solution.

1
Before calibration and deployment, replace the filling solution according to the instructions
below. Repeat filling procedure between deployments or when measurements begin to drift.
Remove sensor from
sonde and unscrew
reference junction.
Squeeze a steady stream
of solution into the
reservoir while slowly
pulling out the tube.
Page 1

Aqua TROLL 500/600 pH/ORP Sensor Overview

The In-Situ pH sensor measures pH levels in natural water, surface water,
groundwater, produced water and aquaculture applications.

Getting Started (3 steps)

Don't pour solution
down the drain. Pour it
onto a paper towel and
discard.
Overfill slightly. Reinstall
reference junction cap
and tighten until it
touches sensor body.
970-498-1500
Instruction Sheet
Lightly shake the bottle
of reference filling
solution to mix.
Turn the cap 90° more
(one quarter of a turn) to
secure.
www.in-situ.com
Insert the fill tube into
the bottom of reservoir.
2022-05-06 | 0063472

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Aqua TROLL 500 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for In-situ Aqua TROLL 500

  • Page 1 Instruction Sheet Aqua TROLL 500/600 pH/ORP Sensor Overview The In-Situ pH sensor measures pH levels in natural water, surface water, groundwater, produced water and aquaculture applications. You must have a Temperature or Conductivity/Temperature sensor installed to use the pH/ORP sensor.
  • Page 2 Connect to the deployment mode after instrument with VuSitu. the menu. option and follow the calibration instructions. For detailed calibration instructions, see the instruction manual or quick start guide for your In-Situ instrument. Page 2 970-498-1500 www.in-situ.com 2022-05-06 | 0063472...
  • Page 3 To remove crystalline water to clean. deposits, clean with warm water and soap, Leave reference solution in the sensor or soak sensor in 5% during storage. HCl solution for 10 to 30 minutes. Page 3 970-498-1500 www.in-situ.com 2022-05-06 | 0063472...

This manual is also suitable for:

Aqua troll 600

Table of Contents