Display Features; Invalid Data Indicator - Keysight Technologies UXA N9040B Getting Started And Troubleshooting Manual

X-series signal analyzer
Hide thumbs Also See for UXA N9040B:
Table of Contents

Advertisement

Front and Rear Panel Features

Display Features

Display Features
This section describes the regions of the display.

Invalid Data Indicator

of this is when instrument settings have changed in the time since the data in
the traces on the display was taken. This means that the screen annotation
cannot be guaranteed to match the trace data. For example, if you change
Center Frequency, the invalid data indicator will display until a new sweep has
completed.
If any Trace is in View mode (displaying but not updating) and analyzer settings
are changed, the invalid data indicator will display as long as that trace
remains in View. Traces that are blanked do not turn on the invalid data
indicator.
Not all analyzer settings require display of the invalid data indicator when they
change, only changes that require a new acquisition will cause it to display.
For example, changing the Y-Axis scale of the analyzer does not cause the
invalid data indicator to display, unless the attenuation changes.
Also, the invalid data indicator is turned on:
— When the counter is turned on, until the completion of the first count
X-Series Signal Analyzers Getting Started and Troubleshooting Guide
The invalid data indicator is displayed whenever the data on the display
does not match the settings of the analyzer. The most common example
25