(13) Disconnect the ESD wrist strap from the ESD connection point inside the enclosure.
(14) Close the enclosure.
(15) Latch the two latches.
(16) If applicable, attach the locking mechanism.
(17) Supply electrical power to the device.

9.1.3 Replace Fuses

The device uses 2 amp, 240 V fuses (MSA P/N 10185821 or an equivalent certified fuse).
Fig. 14
Location of fuses
For a list of approved parts and how to order them, refer to Section 13, "Ordering Information".
When necessary, do the following procedure to replace the fuses inside the devise enclosure:
(1)
Disconnect electrical power to the device.
(2)
If applicable, remove the locking mechanism.
(3)
Release the two latches on the left side of the device.
(4)
Open the enclosure.
(5)
Attach the ESD wrist strap to the ESD connection point inside the enclosure.
(6)
Turn the top of the fuse holder counterclockwise and remove it.
(7)
Remove the defective fuse.
(8)
Install a new fuse. Only use MSA P/N 10185821 or an equivalent certified fuse.
(9)
Install the fuse holder and turn the top clockwise to secure it.
(10) Disconnect the ESD wrist strap from the ESD connection point inside the enclosure.
(11) Close the enclosure.
(12) Latch the two latches.
(13) If applicable, attach the locking mechanism.
(14) Supply electrical power to the device.
Chillgard 5000
Maintenance
US
67