Page 7
Table of Contents Interferant Data ....5-6 Replacement Parts ....5-7 Special Instructions.
Page 8
Table of Contents Section 7 Suggested Daily Use for All Units (MiniCO unit shown) ... . . 7-1 TOC-6 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Section 1, MiniCO Responder Section 1 ® MiniCO Responder Carbon Monoxide Detector (P/Ns 710424, 710510, 10008671 and 10008673) Instructions for Use and Maintenance " WARNINGS FAILURE TO FOLLOW CAN RESULT IN SERIOUS PERSONAL INJURY OR DEATH: 1. The MiniCO Responder is designed to measure carbon monoxide in air only.
Section 1, MiniCO Responder " CAUTIONS To prevent sensor damage, store the instrument under the following conditions: OPTIMAL INTERMITTENT STORAGE TEMPERATURE TEMPERATURE C to 50 C to 32 (-20 F to 120 F to 90 Turning ON/OFF When unit is delivered, the battery is not connected.
Section 1, MiniCO Responder 1/2-second on P/N 710510. The high level alarms are high rate intermittent and latching. This is indicated by an "A" on the display and a beep approximately every 1/2-second. NOTE: On MiniCO Responder (P/Ns 710510 and 10008673), the "A" alarm can be changed to non-latching by setting the alarm setpoint at 999 ppm.
Section 1, MiniCO Responder Calibration Check/Adjustment Turn the unit ON and press and hold both the Zero and Reset; then, Press the Power Peak button. The descriptor "↑SET↓" will appear on the display. The monitor will then automatically zero out the background concentration (display shows "Zero").
Section 1, MiniCO Responder Installing/Replacing the Battery Remove belt clip by pressing the two tabs near the top and releasing from the back. Remove the two slotted head screws on the back of the case, located under the clip. Remove the 9-V battery from the snap connector and replace with a fresh 9-V battery or install a fresh 9-V battery for first time use.
Section 1, MiniCO Responder • The calibration cap is fitted on the instrument, over top of the sensor housing. • Maximum sampling line length is 25 feet. • Minimum sampling line length is 10 feet. Specifications 0-999 ppm, RANGE over range indicated by - - - RESPONSE TIME 90 seconds or less to 90% of final reading TEMPERATURE...
Section 1, MiniCO Responder Interferant Data This data is presented as the indicated output in ppm, which would result from the application of 100 ppm of the test gas. TEST GAS EQUIVALENT PPM Carbon monoxide (CO) 100 +0 Hydrogen sulfide (H Chlorine (CL 3 +1 Nitrogen Dioxide (NO...
Section 2, MiniH S Responder Section 2 ® MiniH Responder Hydrogen Sulfide Detector (P/Ns 710850 and 10008672) Instructions for Use and Maintenance " WARNINGS FAILURE TO FOLLOW CAN RESULT IN SERIOUS PERSONAL INJURY OR DEATH. 1. The MiniH S Responder is designed to measure Hydrogen Sulfide in air only.
Section 2, MiniH S Responder " CAUTIONS To prevent sensor damage, store the instrument under the following conditions: OPTIMAL INTERMITTENT STORAGE TEMPERATURE TEMPERATURE C to 50 C to 32 (-20 F to 120 F to 90 Turning ON/OFF When unit is delivered, the battery is not connected.
Section 2, MiniH S Responder To mute a Warning level, press the RESET button. This silences the audible tone, but the visual indicator remains. If a warning condition remains, the tone will only be temporarily silenced. It is not possible to mute the alarm level while it persists. The alarm level can only be reset once the concentration falls below the alarm level.
Section 2, MiniH S Responder Connect the calibration cap to the sensor cap. Attach the calibration gas to the cal cap. Turn ON the gas. Wait for the display reading to stabilize; then, adjust the concentration on the display by pressing the ↓ (labeled Zero) or ↑ (labeled Reset) to read the same as the concentration of the cylinder of gas attached.
Section 2, MiniH S Responder Replacing Sensor Assembly Remove phillips screw from sensor assembly. Rotate approximately 15 clockwise. Pull assembly out. Remove the shorting bar from the new assembly. Place the assembly in the instrument and rotate approximately 15 counterclockwise; insert the phillips screw and tighten.
Section 2, MiniH S Responder Interferant Data This data is presented as the indicated output in ppm, which would result from the application of 2 ppm of the test gas. EQUIVALEN TEST GAS T PPM Hydrogen Sulfide (H 100 +4 Hydrogen Chloride (HCL) 0 +0 Ethanol (EtOH)
Section 3, MiniOX Responder Section 3 ® MiniOX Responder Oxygen Detector (P/Ns 710965 and 10008676) Instructions for Use and Maintenance " WARNINGS FAILURE TO FOLLOW CAN RESULT IN SERIOUS PERSONAL INJURY OR DEATH. 1. Alarm functions must be checked and a response check must be performed daily.
Section 3, MiniOX Responder " CAUTIONS 1. To prevent sensor damage, store the instrument under the following conditions: OPTIMAL INTERMITTENT STORAGE TEMPERATURE TEMPERATURE C to 50 C to 32 (-20 F to 120 F to 90 2. Acid gases, such as CO , will shorten the life of the sensor.
Section 3, MiniOX Responder The low alarm, indicating oxygen deficiency, is high-rate intermittent and latching. This is indicated by an "A" on the display and a beep approximately every 1/2-second. It is not possible to mute the low alarm level while the deficiency persists.
Section 3, MiniOX Responder appear on the display. The monitor then spans the instrument. When calibration is complete, the instrument beeps three times. Within the next five seconds, if you press and hold the Power Peak button for two seconds, you will be able to change the alarm settings.
Section 3, MiniOX Responder Specifications 0-100% oxygen, RANGE over range indicated by - - - 30 seconds or less RESPONSE TIME to 90% of final reading TEMPERATURE RANGE C to 40 C (14 F to 104 (COMPENSATED) HUMIDITY 15 - 90% (non-condensing) 60 days with five minutes of alarm per day BATTERY LIFE...
Section 4, MiniOX Responder Remote Oxygen Detector Section 4 ® MiniOX Responder Remote Oxygen Detector (P/Ns 10001923 and 10008675) Instructions for Use and Maintenance " WARNINGS FAILURE TO FOLLOW CAN RESULT IN SERIOUS PERSONAL INJURY OR DEATH 1. Alarm functions must be checked and a response check must be performed daily.
Section 4, MiniOX Responder Remote Oxygen Detector " CAUTIONS 1. To prevent sensor damage, store the instrument under the following conditions: OPTIMAL INTERMITTENT STORAGE TEMPERATURE TEMPERATURE C to 50 C to 32 (-20 F to 120 F to 90 2. Acid gases, such as CO , will shorten the life of the sensor Turning ON/OFF...
Section 4, MiniOX Responder Remote Oxygen Detector If a warning condition remains, the tone will only be temporarily silenced. The default setting for the "W" alarm is 22% O The low alarm, indicating oxygen deficiency, is high-rate intermittent and latching. This is indicated by an "A"...
Section 4, MiniOX Responder Remote Oxygen Detector Calibration Check While the instrument is turned ON, press and hold both the Zero and Reset buttons; then, press the Power Peak button. The instrument beeps once and the descriptor "↑SET↓" appears on the display. The monitor then spans the instrument.
Section 4, MiniOX Responder Remote Oxygen Detector Replacing Sensor Assembly Remove the remote oxygen sensor and cable assembly from the instrument by unscrewing the jack assembly counterclockwise and removing jack from the top of the sensor cap. Replace the new oxygen sensor and cable assembly by reversing the above procedure.
Section 5, MiniCl Responder Section 5 MiniCl Responder Chlorine Detector (P/Ns 10020744 and 10020747) Instructions for Use and Maintenance " WARNINGS FAILURE TO FOLLOW CAN RESULT IN SERIOUS PERSONAL INJURY OR DEATH. 1. The MiniCl Responder is designed to measure Chlorine in air only. Do not use this monitor to sample for Cl in gases other than air.
Section 5, MiniCl Responder " CAUTIONS 1. To prevent sensor damage, store the instrument under the following conditions: OPTIMAL INTERMITTENT STORAGE TEMPERATURE TEMPERATURE C to 50 C to 32 (-20 F to 120 F to 90 Turning ON/OFF When unit is delivered, the battery is not connected.
Section 5, MiniCl Responder To mute a Warning level, press the RESET button. This silences the audible tone, but the visual indicator remains. If a warning condition remains, the tone will only be temporarily silenced. It is not possible to mute the alarm level while it persists. The alarm level can only be reset once the concentration falls below the alarm level.
Section 5, MiniCl Responder Connect the calibration cap to the sensor cap. Attach the calibration gas to the cal cap. Turn ON the gas. Wait for the display reading to stabilize; then, adjust the concentration on the display by pressing the ↓ (labeled Zero) or ↑ (labeled Reset) to read the same as the concentration of the cylinder of gas attached.
Section 5, MiniCl Responder has been out of service or the battery has been out of the instrument for more than one day, it is recommended that the instrument not be used or zeroed for a period of 12 hours. This allows the sensor time to equilibrate.
Section 5, MiniCl Responder Specifications DIMENSIONS 4-1/8"(L) x 2-1/4"(W) x 1-3/4" (H) WEIGHT 7 oz. +0.1 ppm Cl at constant temperature; at -10 C and 50 ACCURACY the reading shall not change by more than 0.2 ppm than at ambient temperatures dB LEVEL OF 93 dB at one foot typical ALARMS...
Section 5, MiniCl Responder Replacement Parts COMPONENT/ASSEMBLY PART NO. Sensor Assembly 10021355 9-V Battery 628817 Calibration Cap 10028451 Battery Cover Door with slotted 710493 screws Slotted Screwdriver 632655 Belt Clip 710489 Calibration Gas 710331 Calibration Regulator 465895 Calibration Tubing, 30-inch 485030 Special Instructions Due to the high reactivity of Chlorine gas,...
Section 6, MiniClO Responder Section 6 MiniClO Responder Chlorine Dioxide Detector (P/Ns 10020745 and 10020749) Instructions for Use and Maintenance " WARNINGS FAILURE TO FOLLOW CAN RESULT IN SERIOUS PERSONAL INJURY OR DEATH. 1. The MiniClO Responder is designed to measure Chlorine Dioxide in air only.
Section 6, MiniClO Responder " CAUTIONS 1. To prevent sensor damage, store the instrument under the following conditions: OPTIMAL INTERMITTENT STORAGE TEMPERATURE TEMPERATURE C to 50 C to 32 (-20 F to 120 F to 90 Turning ON/OFF When unit is delivered, the battery is not connected.
Section 6, MiniClO Responder To mute a Warning level, press the RESET button. This silences the audible tone, but the visual indicator remains. If a warning condition remains, the tone will only be temporarily silenced. It is not possible to mute the alarm level while it persists. The alarm level can only be reset once the concentration falls below the alarm level.
Section 6, MiniClO Responder Connect the calibration cap to the sensor cap. Attach the calibration gas to the cal cap. Turn ON the gas. Wait for the display reading to stabilize; then, adjust the concentration on the display by pressing the ↓ (labeled Zero) or ↑ (labeled Reset) to read the same as the concentration of the cylinder of gas attached.
Section 6, MiniClO Responder entered by the same method. After accepting the alarm level, the monitor returns to displaying the concentration reading. Installing/Replacing the Battery Remove belt clip by pressing the two tabs near the top and releasing from the back. Remove the two slotted head screws on the back of the case, located under the clip.
Section 6, MiniClO Responder Specifications 0-6 ppm, RANGE over range indicated by - - - 90 seconds or less RESPONSE TIME to 90% of final reading C to 40 C (14 F to 104 TEMPERATURE C to 50 C if calibrated at RANGE temperature of use and recalibrated again if...
Section 6, MiniClO Responder Interferant Data This data is presented as the indicated output in ppm, which would result from the application of 2 ppm of the test gas. CONCEN- READ- TEST GAS TRATION Methane (CH 1.50% Carbon monoxide (CO) 300 ppm Hydrogen Sulfide (H S) 10 ppm Sulfur Dioxide (SO...
Page 46
Section 6, MiniClO Responder 1. If condensation in the sample line is suspected, dry the sample line by drawing a sample with the sample line attached in a low humidity atmosphere. 2. To verify operation of a MiniClO Responder equipped with a Chlorine Dioxide sensor, perform response check with the sample line in place.
Page 47
Section 7, Suggested Daily Use for All Units Section 7 Suggested Daily Use for All Units (MiniCO unit shown) Turn ON the MiniCO Responder. The instrument can be zeroed in fresh air by pressing the ZERO button while "zero" flashes on the display. Artisan Technology Group - Quality Instrumentation ...
Page 48
Section 7, Suggested Daily Use for All Units Test the instrument’s response to gas: ® Bump check with Squirt Bump Test Gas OR (see step 4): OR: Test the instrument’s response to gas with a calibration gas cylinder. Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 49
Section 7, Suggested Daily Use for All Units The instrument should read within the tolerance printed on the cylinder or can. If not, calibration is required. See "Calibration Check Adjustment." Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Need help?
Do you have a question about the MiniOX 710965 and is the answer not in the manual?
Questions and answers