Sample And Exhaust Lines; Tubing Material - MSA Chillgard 5000 Operating Manual

Refrigerant monitor
Hide thumbs Also See for Chillgard 5000:
Table of Contents

Advertisement

Fig. 8
Front of device
4.4

Sample and Exhaust Lines

4.4.1 Tubing Material

The device can use hard and soft tubing. The following are acceptable tubing types:
US English System
¼ in. OD x 1/8 in. inside diameter (ID) Polyurethane tubing (ether based)
¼ in. OD x 3/16 in. ID Polyurethane tubing (ether based)
¼ in. OD Copper tubing
¼ in. OD Stainless steel tubing
SI Units
6 mm OD x 4 mm ID Polyurethane tubing (ether based)
6 mm OD Copper tubing
6 mm OD Stainless steel tubing
NOTICE
Use the appropriate tools to give a clean, smooth end to the tubing when it is cut.
Fig. 9
Rear of device
Chillgard 5000
Installation
US
18

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents