MSA Chillgard 5000 Operating Manual page 32

Refrigerant monitor
Hide thumbs Also See for Chillgard 5000:
Table of Contents

Advertisement

The default setting for the internal fault relay is the ener-
gized state. To test the relay, do the following:
Select "ON".
Select "OFF".
(2)
In the relay configuration pop-up, select the "Name"
field, and use the alphanumeric keyboard pop-up to
enter a value.
Relay names can have eight characters.
The title in the pop-up header changes to show the
assigned name.
(3)
Select "Energized" or "De-energized" for the relay
state.
(4)
Do the following to test the relay:
a) Select "ON".
b) Make sure the equipment connected to the relay
operates as expected.
c) Select "OFF".
(5)
Select "Save". The updated settings for the relays
show on the "Configure Relays" screen.
(6)
Repeat Steps (1) through (5) for each relay to be
configured.
(7)
When all relays have been configured, select "Next".
Chillgard 5000
Initial Setup