Replacing the Aquathin DI Module
Refer to
(Figure
17) or to page 49, Appendix C2, for
details on releasing tubing from Speedfit® fittings)
Figure 17: Speedfit Push-in Fittings
1.
Disconnect the tubing from each end of the
Aquathin De-ionization (DI) Module.
2.
With
the
DI
module free of
its connections,
it can now be
removed from
t h e
t w o
"Double
C"
clamps.
(Figure
18). Note the
direction of the
" F L O W "
arrow.
The
new DI module must have the same orientation.
3.
Place the new Aquathin DI Module into the
housing. Re-insert the tubing which connects the
reservoir to the DI Module, taking care to attach the
tubing to the outlet end of the module.
4.
Follow the system restarting procedure on page 37.
34
© 2009-2016 Aquathin Corporation. All rights reserved.
Figure 18: Step 2 - removing
the DI module from the clamps