EMC declaration for M300+

General information
This device was tested for electromagnetic compatibility using accessories from the list of accessories.
Other accessories may only be used if they do not compromise the electromagnetic compatibility. The use
of non-compliant accessories may result in increased electromagnetic emissions or decreased
electromagnetic immunity of the device.
This device may be used in the direct vicinity of other devices only if Dräger has approved this device
arrangement. If no approval has been given by Dräger, it must be ensured that this device functions
correctly in the desired arrangement before use. The instructions for use for the other devices must be
followed.
Electromagnetic environment
This device may only be used in environments specified in section "Environment of Use" on page 8.
Emissions
Radiated emissions
Conducted emissions
NOTE
The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals
(CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally
required), this equipment might not offer adequate protection to radio-frequency communication ser-
vices. The user might need to take mitigation measures, such as relocating or re-orienting the equip-
ment.
Immunity against
Electrostatic discharge (ESD) (IEC 61000-4-2)
Magnetic fields at mains frequency
(IEC 61000-4-8)
Radiated high-frequency disturbances
(IEC 61000-4-3)
Conducted high-frequency disturbances
(IEC 61000-4-6)
Electromagnetic fields in the vicinity of wireless
communication devices
Supplement – Infinity M300 and M300+ Series – VG2.4
Electromagnetic compatibility (EMC)
Compliance
Class A, group 1 (30 MHz to 1 GHz)
Class A, group 1 (150 kHz to 30 MHz)
Test level and required electromagnetic environ-
ment
Contact discharge: ±8 kV
Air discharge: ±15 kV
30 A/m
80 MHz to 2.7 GHz: 3 V/m
150 kHz to 80 MHz: 3 V
Various frequencies from 385 MHz to 5785 MHz:
9 V/m to 28 V/m, with various pulse modulations and
immunity test levels.
, ISM bands: 6 V