How To Minimize The Offset Current - Keysight CX3322A Quick Operation Manual

Cx3300 series device current waveform analyzer 2-channel model 4-channel model
Hide thumbs Also See for CX3322A:
Table of Contents

Advertisement

Keysight Technologies
CX3300 Current Waveform Analyzer
Quick Operation Guide

How to minimize the offset current

1.
Perform "User Calibration". The
offset current of the current
sensor will be minimized.
2.
Perform a actual measurement
and fix the connection and the
CX3300 settings such as the
current range.
3.
Perform a "Null" measurement.
3-1. Make a connection to prevent the current flow into the current sensor.
Note for connection:
- Connect either the (+) or (-) terminal of the current sensor to the stable voltage
on the DUT.
- Connect the ground lead to the circuit common of the DUT.
Connection Example
Actual measurement
Ground lead
DUT
VIN
GND
Mainframe
+ Current Sensor
User Calibration
Null measurement
Remove the test lead
of the (+) terminal to
prevent the current
flow into the current
sensor
18
DUT
Null
Ground lead
DUT
VIN
GND

Advertisement