160
•
Figure 9-5: Timing Diagram: Dual-Trigger (Shift Per Trigger) DIF
TRIGGER
IMAGES
9.3.2
Trigger Source
Trigger(s) can either be internally generated by the PI-MAX4 or can be generated by an
external source connected to the
Valid values for
•
•
9.3.2.1
Internal Trigger Source
Trigger pulses are generated by the PI-MAX4 using the configured
Frequency
Figure 9-6: SuperSYNCHRO Timing Expander: DIF Operation
The range of valid frequencies is 1 Hz to 1 MHz, in 1 Hz increments.
The
Internal Trigger Frequency
and
SyncMASTER2
®
PI-MAX
Shift Per Trigger
Two triggers are used to acquire the two images (i.e., one trigger per image.)
Figure 9-5
illustrates the timing diagram for a dual-trigger DIF acquisition.
FIRST FRAME
D1
W1
TT
PD: Phosphor Decay Time
W1: Pulse 1 Gate Width
TRIGGER IN
Trigger Source
;
Internal
External
.
on the
SuperSYNCHRO Timing
outputs.
4 System Manual
SHIFT
BEHIND
MASK
PD
S)
D2: Pulse 2 Gate Delay
W2: Pulse 2 Gate Width
connector on the rear of the camera.
are:
expander. See
setting also determines the frequency of
SECOND FRAME
D2
W2
PD
Internal Trigger
Figure
9-6.
SyncMASTER1
Issue 9
Need help?