Short Form Instructions - Welch Allyn AT-1 Manual

3-channel electrocardiograph
Table of Contents

Advertisement

AT-1 -
User Guide

Short Form Instructions

Automatic ECG Recording
Prepare skin, hook up patient
Press ON
Wait for at least 10 seconds for the ECG to stabilise.
Press AUTO
Press COPY
Press OFF
to switch unit on.
to record and print.
for additional copies.
to switch unit off.
Manual ECG Recording (Rhythm Strip)
Prepare skin, hook up patient
Press ON
Press MAN START
Change lead group with
Press STOP
Electrode hook-up check
Press key sequence
The contact resistance of each electrode is printed. Best results are
obtained when the electrode voltage readings (right column) are between
±50mV.
Filter On/Off
Press
to switch the (Myogram) filter On / Off.
Print System Configuration
Press key sequence
to switch unit on.
to start printout.
and
.
to stop the printout.
AT-1 User Guide Art. No. 2. 510 270
.
.
1