Power; Gas Supplies - Agilent Technologies VGA 77 User Manual

Vapor generation accessory
Table of Contents

Advertisement

Condition
Non-operating (transport)
Operating within performance
specifications

Power

If necessary, replace the power cord only with a cord equivalent to
the one specified below.
Information is correct at time of printing—information on the back of
the accessory is the most up-to-date.
Voltage
Frequency
Consumption
Connections
Mains inlet coupler: 6 A 250 VAC IEC type
Mains power cord connector:
Australia
USA
Europe

Gas Supplies

Argon or nitrogen, 99.99% pure (minimum)
Only argon gas can be used when the VGA 77 is coupled with an ICP
NOTE
spectrometer.
The inert gas must be dry and dust-free, otherwise you must insert a
suitable filter in the supply line.
Agilent Vapor Generation Accessory VGA 77 and VGA 77P User's Guide
Altitude
0-2133 m
(0-7000 ft)
0-853 m (0-2800 ft)
853-2133m (2800-7000 ft)
100, 120, 220, or 240 volts AC
50-60 Hz
approx. 20 VA
10 A 250 VAC
Complies with AS3112
10 A 125 VAC
Complies with NEMA 5-15P
6 A 250 VAC
Complies with CEE7 sheet vii or NFC61.303VA
Humidity (%RH)
Temp (°C)
non-condensing
5-45
20-80
10-35
8-80
10-25