Table 3: Reference Values for Stimulus Calibration
 
Freq.
Reference values for stimulus calibration
[dB re. 20 µPa]
   
   
 
EARtone
[Hz]
TDH39
 
 
3A / ABR
 
125
45
26
 
 
250
25.5
14
 
 
500
11.5
5.5
 
 
1000
7
0
 
 
1500
6.5
2
 
 
2000
9
3
 
 
3000
10
3.5
 
 
4000
9.5
5.5
 
 
6000
15.5
2
 
 
8000
13
0
 
 
WB
 
-8
-5
LP
 
-6
-7
HP
 
-10
-8
CE-
27.5
31.5
Chirp
CE-
26.5
26.5
Low
Chirp
CE-
28
31
High
Chirp
Click
30.5
35
*All figures in bold are Interacoustics Standard values.
Titan
 
CIR55
DD45 IOW
 
26
 
47.5 41
14
 
27
5.5
 
13
0
 
6
2
 
8
3
 
8
3.5
 
8
5.5
 
9
2
 
20.5 3
0
 
12
 
-5
-8
 
-7
-6
 
-8
-10
26
25.5 27.5
28
32.5 28
Setup and Installation Manual
IOW
NB
Probe
Stimulus
Probe
NB
Correction
 
Values
(except IOW
probe)
4
43.5
4
24.5
26.5
4
9.5
17
6
6.5
10.5
6
5
12
6
12
11
6
11
11
5
3.5
8
5
5.5
5
-5
-0.5
-5
 
 
-7
 
 
-8
 
 
32
32
Variation of Ipsi
stimulus levels
for different
volumes of the
 
ear canal
Relative to the
calibration
performed on an
IEC 126 coupler
[dB]
 
 
 
0.5 ml
1 ml
 
 
 
 
 
 
9.7
5.3
 
9.7
5.3
 
 
 
11.7
3.9
 
-0.8
-0.5
 
-1.6
-0.8
 
 
 
 
 
7.5
3.2
 
8.0
3.6
 
3.9
1.4
 
Page 37
Sound
attenuation
values for
TDH39
earphones
using
MX41/AR or
PN51
 
cushion
[dB]
 
3
 
5
 
7
 
15
21 (1600 Hz)
26
31 (3150 Hz)
32
26 (6300 Hz)
24
 
 
 
 
 
 
 

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Titan dpoae440Titan abris440

Table of Contents