Reflex Functions
Signal sources:
Outputs:
Test types:
ABR Infant Screening  
Preamplifier
Electrical Impedance
measurement
Stimulus
Recording
Display
Algorithmic Sensitivity
Algorithmic Specificity
Titan
Tone - Contra, Reflex:
Tone - Ipsi, Reflex:
NB noise - Contra, Reflex:
NB noise - Ipsi, Reflex:
Noise - Contra, Reflex:
Noise - Ipsi, Reflex:
Contra Earphone:
Ipsi Earphone:
Air:
Manual Reflex
Automated Reflex
Reflex Decay
Reflex latency
One Channel
Gain:
Frequency response:
Noise:
CMR Ratio:
Max input offset voltage:
Input impedance:
Power from main unit:
Measurement frequency:
Waveform:
Measurement current:
Range:
Stimuli
Stimulus rate:
Transducers:
Channels:
Level:
Bandwidth:
Analysis time:
A/D resolution:
Artifact reject system:
CE-Chirp
CE-Chirp
Setup and Installation Manual
250, 500, 1000, 2000, 3000, 4000, 6000, 8000 Hz.
500, 1000, 2000, 3000, 4000 Hz.
250, 500, 1000, 2000, 3000, 4000, 6000, 8000 Hz.
1000, 2000, 3000, 4000 Hz.
Wide Band, High Pass, Low Pass.
Wide Band, High Pass, Low Pass.
TDH39 earphone, DD45 earphone, CIR55 insert and/or EARtone 3A
insert for Reflex measurements.
Probe earphone incorporated in the probe system for Reflex
measurements.
Connection of the air system to the probe.
Manual control of all functions.
Automatic reflexes:
-
Single intensities
-
Reflex growth
Automatic, 10 dB above threshold and manually controlled with
stimulus durations of 10 to 30 s.
Automated, first 300 ms from stimulus start.
3 electrodes. 50 cm
Switchable: Software will automatically switch mastoid and ground if
mastoid montage is used. So the user does not need to replace
electrode during testing.
64 dB
0,5 - 5000 Hz
<25 nV/√Hz
>90 dB.
2.5 V
10 MΩ/ 170 pF
Isolated power supply
33 Hz
Rectangular
11.25 µA
0.5 kΩ – 25 kΩ ± 10 %
Click range (177-11313 Hz)
CE-Chirp® range (177 – 11313 Hz)
HiLo CE-Chirp® range
(Lo 177- 1414 Hz) & (Hi 1414 – 11313 Hz)
90 Hz
EarTone ABR insert phone auto calibrated & detection TDH 39 or
DD45 head phone auto calibrated & detection
IOW Probe auto calibrated & detection
2
30 dB nHL, 35 dB nHL, 40 dB nHL
22.05 kHz
1-10 min or Residual noise 5-80 nV
24 bit
Rejection level (Peak, Min RMS, Max RMS) Clipping (Saturation)
Stimulus level and type, Bar and Graph view
Basic and advanced view
99.9%
> 96%