ct through cradle
up to 80%
Off
On (not testing)
2.2.1.11

Changing Probes and Extension Cords on the Titan

The Titan probe can be connected in four ways to the device:
1) directly to the Titan
2) using the short extension cord
3) using the long extension cord with the shoulder box (small – for OAE & IMP)
4) using the long extension cord with the PreAmp (large – for OAE/IMP & ABRIS)
Note that the length of the tubing influences the pump speed. When the probe is directly connected to the
Titan, pump speeds are about 60% higher than when it is connected via the long extension cable with either
the shoulder box or PreAmp. For the most accurate measurements it is recommended to use probe while
connected via the shoulder box or PreAmp.
Titan
ct through USB
(PC) up to 80%
1,5
3,8
2,8
-32
To release the probe or the extension cable from the Titan,
press the button at the back of the Titan and pull the probe or
extension cord out while holding onto the base of the probe or
extension cable.
Do not pull directly on the cord of the extension cable as this
can damage the tubing connections! Do not pull directly on
the probe tip as this may cause it to break!
Connect the extension cord or probe to the Titan by lining up
the red triangles and pushing the probe or extension cable
into the connection plate.
The probe can be attached to the extension cable by lining it
up correctly and clicking the probe onto the end. .
Setup and Installation Manual
ct through cradle
ct through USB
up to 100%
(PC) up to 100%
2,3
4,1
5,7