Ethernet Settings Menu - Ametek 7270 Instruction Manual

Dsp lock-in amplifier
Table of Contents

Advertisement

Chapter 5, FRONT PANEL OPERATION

5.3.18 Ethernet Settings Menu

5-32
instrument manually from a simple RS232 terminal where the maximum speed
with which characters can be sent to the instrument is limited by the speed of
human typing.
PROMPT
This control has two settings, as follows:-
ON
A prompt character is generated by the model 7270 after each command
response to indicate that the instrument is ready for a new command. The prompt
character is either a "*" or a "?" If a "?" is generated, it indicates that an
overload, reference unlock, parameter error or command error has occurred.
OFF
No prompt character is generated.
DELIMITER
The character shown is that sent by the lock-in amplifier to separate two numeric
values in a two-value response, such as that generated by the MP command. The
corresponding ASCII value of this character is also shown in brackets. For example,
value 44 corresponds to a "," (comma).
PARITY
This control sets the parity check polarity when the Data Bits control specifies that a
parity bit should be used. It should be set to match the setting of the controlling
computer.
Pressing the Previous Menu key returns control to the Communications Menu, and
pressing the Communications Monitor key accesses the Communications Monitor
display.
When the Communications menu is displayed, pressing a key adjacent to the
Ethernet Settings item accesses the Ethernet Settings menu, which is shown in figure
5-23.
This menu has one control, five keys to access other menus, and four indicators for
Figure 5-23, Ethernet Settings Menu