2.2 Initial Checks

2.2.01 Introduction

2.2.02 Procedure

The following procedure checks the performance of the model 7270. In general, this
procedure should be carried out after inspecting the instrument for obvious shipping
damage.
NOTE: Any damage must be reported to the carrier and to SIGNAL RECOVERY
immediately. In addition the shipping container must be retained for inspection by
the carrier.
Note that this procedure is intended to demonstrate that the instrument has arrived in
good working order, not that it meets specifications. Each instrument receives a
careful and thorough checkout before leaving the factory, and normally, if no
shipping damage has occurred, will perform within the limits of the quoted
specifications. If any problems are encountered in carrying out these checks, contact
SIGNAL RECOVERY or the nearest authorized representative for assistance.
1) Ensure that the model 7270 is set to the line voltage of the power source to be
used, as described in section 2.1.05.
2) With the rear-panel mounted power switch set to 0 (off), plug in the line cord to
an appropriate line power source.
3) Turn the model 7270 power switch to the I (on) position.
4) The front panel display will now briefly display the following:-
5) Wait until the opening display has changed to the Main Display and then press
the key under the bottom right hand corner of the display identified by the legend
MENU on the display. This enters the first of the two main menus,
Main Menu 1, shown below in figure 2-3.
Chapter 2, INSTALLATION AND INITIAL CHECKS
Figure 2-2, Opening Display
2-3