Ametek 7270 Instruction Manual page 159

Dsp lock-in amplifier
Table of Contents

Advertisement

SYNCOSC [n]
Synchronous Oscillator control
When external reference mode is active, the parameter n sets the synchronous
oscillator output in accordance with the following table:
n
Function
0
The oscillator is independent of the external reference frequency
1
The oscillator is a sine wave at the same frequency as the applied reference
signal but subject to any phase shift set by the phase control.
ASTART[.] [n]
Oscillator amplitude sweep start amplitude
Sets the start amplitude for a subsequent sweep of the internal oscillator amplitude,
in the range 0 to 5.000 V
In fixed point mode, n is in microvolts rms, and in floating point mode n is in volts
rms
ASTOP[.] [n]
Oscillator amplitude sweep stop amplitude
Sets the stop amplitude for a subsequent sweep of the internal oscillator amplitude,
in the range 0 to 5.000 V
In fixed point mode, n is in microvolts rms and in floating point mode n is in volts
rms
ASTEP[.] [n]
Oscillator amplitude sweep step size
Sets the amplitude step size for a subsequent sweep of the internal oscillator
amplitude, in the range 0 to 5.000 V
In fixed point mode, n is in microvolts rms and in floating point mode n is in volts
rms
OF[.] [n]
Oscillator frequency control
In fixed point mode n sets the oscillator frequency in mHz. The range of n is 0 to
250,000,000 representing 0 to 250 kHz.
In floating point mode n sets the oscillator frequency in Hz. The range of n is 0 to
2.5E5
FSTART[.] [n]
Oscillator frequency sweep start frequency
Sets the start frequency for a subsequent sweep of the internal oscillator frequency,
in the range 0 to 250 kHz.
In fixed point mode, n is in millihertz, and in floating point mode n is in hertz.
FSTOP[.] [n]
Oscillator frequency sweep stop frequency
Sets the stop frequency for a subsequent sweep of the internal oscillator frequency,
in the range 0 to 250 kHz.
In fixed point mode, n is in millihertz and in floating point mode n is in hertz.
FSTEP[.] [n
n
]
Oscillator frequency sweep step size and type
1
2
The frequency may be swept either linearly or logarithmically, as specified by
parameter n
. The step size is specified by parameter n
2
Log sweep n
= 0
2
In fixed point mode, n
Chapter 6, COMPUTER OPERATION
is the step size in thousandths of a percent.
1
.
1
6-21