6.5 Programming Examples

6.5.01 Introduction

6.5.02 Basic Signal Recovery

X[.]
XOF[n
[n
]]
1
2
XY[.]
Y[.]
YOF[n
[n
]]
1
2
AQN
AS
AXO
This section gives some examples of the commands that need to be sent to the lock-
in amplifier for typical experimental situations.
In a typical simple experiment, the computer is used to set the instrument controls
and then to record the chosen outputs, perhaps as a function of time. At sampling
rates of up to a few points per second, there is no need to use the internal curve
buffer. The commands to achieve this would therefore be similar to the following
sequence:
IE 2
Set reference to external front panel input
VMODE 1
Single-ended voltage input mode
AUTOMATIC 1
AC Gain control automatic
FLOAT 1
Float input connector shell using 1 k to ground
LF 0 0
Turn off line frequency rejection filter
ASM
Auto-Measure (assumes reference frequency > 1 Hz)
TC 13
Set time constant to 200 ms, since previous ASM changed it
Then the outputs could be read as follows:
X.
Reads X channel output in volts
Y.
Reads Y channel output in volts
MAG.
Reads Magnitude in volts
PHA.
Reads Phase in degrees
FRQ.
Reads reference frequency in hertz
The controlling program would send a new output command each time a new reading
were required. Note that when using an output filter slope of 12 dB/octave a good
"rule of thumb" is to wait for a period of four time-constants after the input signal
has changed before recording a new value. Hence in a scanning type experiment, the
program should issue the commands to whatever equipment causes the input signal
to the lock-in amplifier to change, wait for four time-constants, and then record the