LOG Key
Purpose
Use the LOG key to automatically record basic electri-
cal data to the 928A flash memory over a specified time
interval. This function, also called AUTOLOG, makes
it convenient to record measured electrical quantities
while the equipment is unattended.
Logging Features
Basic Data only (see list at end of section).
Logging interval in seconds
Stores up to 6592 records, maximum
Viewing Autolog Information
To view Autolog information, press 2nd
the screen will display some basic information about
the data logging features.
In order to operate, the Autolog function needs
start and stop times and dates and a logging interval in
seconds. Set up these features while in the AUTOLOG
INFO menu by pressing 2nd
Viewing Autolog Data
Use Mlink software (see page 80) to download Autolog
records to computer and view in spreadsheet format.
They cannot be viewed on the 928A itself.
LOG, and
MENU.
62